High Performance Differential Probe Supports 3 Types of Measurements

제조사
HIOKI
제품명
DIFFERENTIAL PROBE 9322
제품번호
9322

견적문의

PRODUCT INFO. 제품상세정보

Measurement functions

DC mode: Waveform monitor output, DC to 10 MHz ±3 dB
AC mode: Detection of power line surge noise, 1 kHz to 10 MHz ±3 dB
RMS mode: Rectified RMS output of DC and AC voltages, DC, 40 Hz to 100 kHz, Response speed: 200 ms or less (400 VAC)

Input type

Balanced differential input

Output

Voltage division ratio: 1/1000, BNC terminal (DC/AC/RMS 3-mode selectable output)

DC amplitude accuracy

±1 % f.s. (1000 VDC or less), ±3 % f.s. (2000 VDC or less) (f.s.=2000 VDC)

RMS amplitude accuracy

±1 % f.s. (DC, 40 Hz to 1 kHz), ±4 % f.s. (1 kHz to 100 kHz) (f.s.=1000 VAC)

Input resistance, capacity

H-L: 9 MΩ, approx 10 pF (C at 100 kHz)
H-case, L-case: 4.5 MΩ, approx 20 pF (C at 100 kHz)

Max. allowable input

600V AC/DC (CAT III), 2000 VDC, 1000 VAC (CAT II)

Max. rated voltage to earth

When using grabber clip: 600 V AC/DC (CAT III), 1500 V AC/DC (CAT II)
When using alligator clip: 600 V AC/DC (CAT III), 1000 V AC/DC (CAT II)

Power supply

1) AC ADAPTER 9418-15 (12 VDC ±10 %) *1
2) Power supply through Power Cord 9324 connected to logic terminal on Memory HiCorder, or other method
*1 Operating voltage range: +5 to +12 V, less than 300 mA. DC jack OD 5.5 mm (0.22 in), ID 2.1 mm (0.08 in)

Dimensions and mass

70 mm (2.76 in) W × 150 mm (5.91 in) H × 25 mm (0.98 in) D, 350 g (12.3 oz)

Accessories

Alligator clips ×2, Grabber clip 9243 ×1 (Red/black each one), Carrying case 3853 ×1, Instruction manual ×1

■ How to power the 9322 with Logic terminal on Memory HiCorder*1 Discontinued model *2 Not including the Logic terminals with the Logic Unit 8973, table indicates the number of 9322 that can be powered from the main unit’s logic terminals

Main unit

Required power cord (s)

Number of Max. connectable 9322s

MR8880-20

Power cannot be supplied from the logic terminals

MR8875

Via the Power Cord 9328 connected to DC output power terminal on the MR8875, up to 3 × 9322
(Note) Power cannot be supplied from the logic terminals

MR8870-20

Power cannot be supplied from the logic terminals

8870-20 *1

Power cannot be supplied from the logic terminals

8861-50 8861 *1

9324 + 9323

2

8860-50 8860 *1

9324 + 9323

2

8855 *1

9324 + 9323

2

8847 *1 MR8847-01 *1 MR8847-02 *1 MR8847-03 *1 MR8827 *2 MR8847-51 *2 MR8847-52 *2 MR8847-53 *2

9324 + 9323

4 *2

8841 *1 8842 *1

9324

1

8835-01 *1 8835 *1

9324

1

8826

9324

4

■ How to power the 9322 with F/V Unit 8940’s sensor terminal*1 Discontinued model *2 Not including the Logic terminals with the Logic Unit 8973, table indicates the number of 9322 that can be powered from the main unit’s logic terminals

Main unit

Required power cord

Number of Max. connectable 9322s

Max. connectable 9322s when simultaneously using clamp sensors

8861-50 8861 *1

9325

6

8

8860-50 8860 *1

9325

6

8

8841 *1 8842 *1

9325

6

4

8835-01 *1 8835 *1

9325

4

4

8826

9325

6

6

■ How to power the 9322 with a Hioki Memory HiCorder 8860series, or 8855*1 Discontinued model *3 Depends on the combination of Clamp-on probes connected to the 9687; number of connectable 9322 are different

Main unit

PROBE POWER UNIT 9687
[For 8860 series]

Input units 8950, 8952, 8953-10, 8955 [For 8855]

 

Use with the Power Cord 9248

Use with the Power Cord 9328

8861-50 8861 *1

8 *3

N/A

8860-50 8860 *1

8 *3

N/A

8855 *1

 

 


PRODUCT INFO. 제품상세정보